Probe Adapters

Quick Links:

Non ZIF Adapters for Commonly Used Electrodes

Channels

Part #

Preparations

Probe Connector

Mates With

16

16

32

32

Chronic

Acute

Acute

Acute

18-pin female Omnetics Nano dual row header

18-pin female DIP socket header (0.5mm pins)

40-pin Samtec FOLC high density socket strip (1.27mm pins)

40-pin Samtec FOLC high density socket strip (1.27mm pins)

16 Channel Acute Headstage to 16 Channel Chronic Probe

Part Number: AC-CH

 

This adapter allows the user to connect a 16-channel chronic probe (such as a TDT Omn1010 microwire array) to a RA16AC or RA16AC4 acute headstage. Standard operation for chronic preparations is single ended with ground and reference shorted together in the chronic headstage. However, the acute headstage is designed for differential operation. When using the acute headstage with our microwire arrays, short G and R together on the adapter for single ended operation.

16 Channel Chronic Headstage to 16 Channel Acute Probe

Part Number: CH-AC

 

This adapter connects a 16-channel acute probe to a chronic headstage (RA16CH). Reference and ground are tied together by default on the chronic headstage so in general only one pin connection is necessary. A jumper is provided on the RA16CH for differential operation.

16 Channel Chronic Headstage to 32 Channel Acute Probe

Part Number: CHx2-NN

 

This adapter allows the user to connect a 16-channel chronic probe (such as a TDT Omn1010 microwire array) to a RA16AC or RA16AC4 acute headstage. Standard operation for chronic preparations is single ended with ground and reference shorted together in the chronic headstage. However, the acute headstage is designed for differential operation. When using the acute headstage with our microwire arrays, short G and R together on the adapter for single ended operation.

16 Channel Acute Headstage to 32 Channel Acute Probe

Part Number: ACx2-NN

 

This adapter connects a 32-channel acute Samtec MOLC probe (compatible with NeuroNexus) to two acute headstages (RA16AC/RA16AC4). Standard operation with the NeuroNexus probe is differential. If you wish to use the Reference pad on the probe, do not tie G and R together.